EasyManua.ls Logo

Anritsu MS2840A - [:SENSe]:MIXer:LOSS[:FIXed] <power>

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 SCPI Device Message Detail
2-996
[:SENSe]:MIXer:LOSS[:FIXed] <power>
External Mixer Conversion Loss Fixed Value
Function
This command sets the external mixer's fixed conversion loss.
Command
[:SENSe]:MIXer:LOSS[:FIXed] <power>
Parameter
<power>
Conversion Loss of External Mixer
Range 0.00 to 99.99 dB
Resolution 0.01 dB
Suffix Code DB
dB is used when omitted.
Default 15.00 dB
Details
This function is available when Option 044/045 is installed for MS2830A.
This function is available when Option 044/046 is installed for MS2840A.
A value can be set per external mixer band.
Example of Use
To set the external mixer's fixed conversion loss to 10 dB.
MIX:LOSS 10DB