Chapter 2 SCPI Device Message Details
2-56
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel <real>
Reference Level
Function
This command sets the reference level.
Command
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel <real>
Parameter
<real>
Reference level
Range Value equivalent to –120 to +50 dBm
Resolution 0.01 dB (When scale unit settings are dB-system
units)
0.01 pV (When scale unit settings are V-system
units)
0.01 yW (When scale unit settings are W-system
units)
Suffix code
DBM,DM
dBm
DBMV
dBmV
DBUV
dBµV
DBUVE
dBµV (emf)
DBUVM
dBµV/m
V
V
MV
mV
UV
µV
W
W
MW
mW
UW
µW
NW
nW
PW
pW
FW
fW
Scale unit settings apply when omitted.
V is used for Linear Scale.
Default value 0 dBm
Details
This command is not available when Spurious Emission is On and Gate