2.9 ACP
2-383
2
SCPI Device Message Details
[:SENSe]:ACPower:FILTer[:RRC][:STATe] OFF|ON|0|1
Adjacent Channel Power Offset Filter Type
Function
This command sets the filter type of the offset channel for Adjacent
Channel Power measurement.
Command
[:SENSe]:ACPower:FILTer[:RRC][:STATe] <switch>
Parameter
<switch>
Filter type
OFF|0
Rectangle filter
ON|1
Root Nyquist filter
Example of Use
To set the filter type of the offset channel for ACP measurement to Root
Nyquist.
ACP:FILT ON
Related command
This command has the same function as the following command.
:CALCulate:ACPower:FILTer[:RRC][:STATe]
:CALCulate:ACPower:FILTer[:RRC][:STATe] OFF|ON|0|1
Adjacent Channel Power Offset Filter Type
Function
This command sets filter type of the offset channel for Adjacent Channel
Power measurement.
Refer to
[:SENSe]:ACPower:FILTer[:RRC][:STATe].
Related command
This command has the same function as the following command.
[:SENSe]:ACPower:FILTer[:RRC][:STATe]