EasyManua.ls Logo

Anritsu MS2840A - CALC:MARK:TRCK;CALC:MARK:TRCK?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter
Native Device Message Details
4-53
4
Native Device Message Details
CALC:MARK:TRCK/CALC:MARK:TRCK?
Marker Tracking
Function
This command sets the Marker tracking function to On/Off.
Command
CALC:MARK:TRCK switch_com
Query
CALC:MARK:TRCK?
Response
switch_res
Parameter
switch_com
Marker tracking function On/Off
ON|1
Marker tracking On
OFF|0
Marker tracking Off
switch_res
Marker tracking function On/Off
1
Marker tracking function On
0
Marker tracking function Off
Details
Conducts sweep after adjusting the center frequency of trace to that of
the active marker frequency.
Example of Use
To enable Marker tracking
CALC:MARK:TRCK ON
CALC:MARK:TRCK?
> 1