Chapter
4
Native Device Message Details
4-53
4
Native Device Message Details
CALC:MARK:TRCK/CALC:MARK:TRCK?
Marker Tracking
Function
This command sets the Marker tracking function to On/Off.
Command
CALC:MARK:TRCK switch_com
Query
CALC:MARK:TRCK?
Response
switch_res
Parameter
switch_com
Marker tracking function On/Off
ON|1
Marker tracking On
OFF|0
Marker tracking Off
switch_res
Marker tracking function On/Off
1
Marker tracking function On
0
Marker tracking function Off
Details
Conducts sweep after adjusting the center frequency of trace to that of
the active marker frequency.
Example of Use
To enable Marker tracking
CALC:MARK:TRCK ON
CALC:MARK:TRCK?
> 1