SCPI Device Message Details
:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10:X:POSition <integer>
Zone Marker Position
Function
This command moves the center of the zone marker to the specified
position.
Command
:CALCulate:MARKer[n]:X:POSition <integer>
Parameter
<n>
Marker number
1
Marker 1
2
Marker 2
3
Marker 3
4
Marker 4
5
Marker 5
6
Marker 6
7
Marker 7
8
Marker 8
9
Marker 9
10
Marker 10
When omitted: Marker 1
<integer>
Number of display points from the left edge
of the screen
Range
0 to 10, 20, 40, 50, 100, 200, 250, 400, 500, 1000,
2000, 5000, 10000, 30000 (MS269xA Only)
Details
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Not available during Spectrum Emission Mask measurement.
Example of Use
To move the center of the zone marker to the 5 hundredth point from the
left edge of the screen.
CALC:MARK:X:POS 500