Chapter 2 SCPI Device Message Details
2-148
:CALCulate:PMARker[1]|2|3|4|5|6|7|8|9|10:Y?
Power Marker Result Query
Function
This command queries the total power and the power density within the
range of the zone marker.
Query
:CALCulate:PMARker[n]:Y?
Response
<power>,<density>
Parameter
<n>
Marker number
1
Marker 1
2
Marker 2
3
Marker 3
4
Marker 4
5
Marker 5
6
Marker 6
7
Marker 7
8
Marker 8
9
Marker 9
10
Marker 10
When omitted: Marker 1
<power>
Total power within the range of zone marker
(/Zone)
<density>
Power density within the range of zone marker
(/Hz)
When scale unit settings are dB-system units.
Resolution 0.001 dB
Suffix code None. Value is returned in the unit specified by
Scale Unit.
–999.0 is returned when no measurement is
performed or an error occurs.
–999.0 is returned when Marker Mode is set to
Fixed or Off.
–999.0 is returned when Zone Width Type is set
to Spot.