2.18 Other Settings
2-955
2
SCPI Device Message Details
Example of Use
To query the title character string.
DISP:ANN:TITL:DATA?
> SPECTRUM ANALYZER
Related Command
This command has the same function as the following command.
:DISPlay:ACPower:ANNotation:TITLe:DATA
:CALibration:YTF
Pre-selector Auto Tune
Function
This command sets the Pre-selector to Auto.
Command
:CALibration:YTF
Details
[MS269xA], [MS2830A], [MS2840A]
This command is not available during the Spurious Emission
measurement and when Displayed Segment Mode is set to Auto.
[MS2830A]
This command is not available for MS2830A-040/041/043.
[MS2840A]
This command is not available for MS2840A-040/041.
Example of Use
To set the pre-selector to auto.
CAL:YTF