Native Device Message Details
4-322
SWE:EGAT:TIME/SWE:EGAT:TIME?
Gate View Sweep Time
Function
This command sets the sweep time for Gate View.
This parameter and the one for setting the sweep time for time axis
measurement reference the same value.
Command
SWE:EGAT:TIME time
Query
SWE:EGAT:TIME?
Response
time
Parameter
time
Sweep time
Range 1
µ
s to 1000 s
Resolution 1 ms (time
1 s)
0.1 s (1 s <time < 10 s)
1 s (10 s
time)
Suffix code NS,US,MS,S
S is used when omitted.
Details
This command is not available when Gate View is set to Off.
Example of Use
To set the sweep time to 100 ms.
SWE:EGAT:TIME 0.1
SWE:EGAT:TIME?
> 0.100000