2
SCPI Device Message Details
[:SENSe]:ACPower:ADVanced:OFFSet:BANDwidth[:INTegration]
<bandwidth>,<bandwidth>,<bandwidth>,<bandwidth>,<bandwidth>,<bandwid
th>,<bandwidth>,<bandwidth>
Adjacent Channel Power Offset Channel Bandwidth
Function
This command sets offset channel bandwidth for Adjacent Channel
Power measurement.
Command
:SENSe]:ACPower:ADVanced:OFFSet:BANDwidth[:INTegration]
<bandwidth_1>,<bandwidth_2>,<bandwidth_3>,<bandwidth_4>,
<bandwidth_5>,<bandwidth_6>,<bandwidth_7>,<bandwidth_8>
Parameter
<bandwidth_n>
Channel bandwidth of Offset_n
Range 1 Hz to 125 MHz
Resolution 1 Hz
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
Default 3.84 MHz
Details
This function can be set when the following trace is active.
•
Spectrum
To query the measurement results after executing this command, use the
*WAI command to control synchronization.
Example of Use
To set the Offset Channel bandwidth to 3.84 MHz and query the results.
ACP:ADV:OFFS:BAND
3.84MHZ,3.84MHZ,3.84MHZ,3.84MHZ,3.84MHZ,3.84MHZ,3.84MHZ,
3.84MHZ
*WAI
FETC:ACP?