2.13 Spectrum Emission Mask
2-591
2
SCPI Device Message Details
:CALCulate:SEMask:CARRier[:POWer]?
Spectrum Emission Mask Reference Power Query
Function
This command queries the reference power for Spectrum Emission Mask
measurement.
Query
:CALCulate:SEMask:CARRier[:POWer]?
Response
<ampl>
Parameter
<ampl>
Reference power
Range –200 to 200 dBm
Resolution 0.01 dB
Suffix code None. Value is returned in dBm units.
Details
This command is not available when Reference Mode is set to either to
the following:
•
Channel
•
Peak
Example of Use
To query the reference power.
CALC:SEM:CARR?
> -10.00
Related command
This command has the same function as the following command.
[:SENSe]:SEMask:CARRier[:POWer]?