2.4 Marker
2-129
2
SCPI Device Message Details
Example of Use
To query the zone marker width by the number of screen segments.
CALC:MARK:WIDT:DIV?
> 5
:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10:WIDTh:POINt <integer>
Zone Marker Width (by Point)
Function
This command sets the zone marker width by the point.
Command
:CALCulate:MARKer[n]:WIDTh:POINt <integer>
Parameter
<n>
Marker number
1
Marker 1
2
Marker 2
3
Marker 3
4
Marker 4
5
Marker 5
6
Marker 6
7
Marker 7
8
Marker 8
9
Marker 9
10
Marker 10
When omitted: Marker 1
<integer>
Zone marker width
Range 1 to 11, 21, 41, 51, 101, 201, 251, 401, 501, 1001,
2001, 5001, 10001, 30001 (MS269xA Only)
points (number of trace points)
Details
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Not available during Spectrum Emission Mask measurement.
Example of Use
To set the zone marker width of Marker 1 to 501 points.
CALC:MARK:WIDT:POIN 501