EasyManua.ls Logo

Anritsu MS2840A - :DISPlay:SEMask:RESult:TYPE PEAK|MARGin; :DISPlay:SEMask:RESult:TYPE?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.13 Spectrum Emission Mask
2-689
2
SCPI Device Message Details
:DISPlay:SEMask:RESult:TYPE PEAK|MARGin
Spectrum Emission Mask Result Type
Function
This command switches the type of the result display for Spectrum
Emission Mask measurement.
Command
DISPlay:SEMask:RESult:TYPE <type>
Parameter
<type>
Type of result display
PEAK
Displays the peak
MARGin
Displays the margin to the limit line.
Example of Use
To set the type of the result display to the peak.
DISP:SEM:RES:TYPE PEAK
:DISPlay:SEMask:RESult:TYPE?
Spectrum Emission Mask Result Type Query
Function
This command queries the type of the result display for Spectrum
Emission Mask measurement.
Command
DISPlay:SEMask:RESult:TYPE?
Response
<type>
Parameter
<type>
Type of result display
PEAK
Displays the peak
MARG
Displays the margin to the limit line.
Example of Use
To query the type of the result display.
DISP:SEM:RES:TYPE?
> PEAK