2.7 Sweep/Trigger/Gate
2-309
2
SCPI Device Message Details
Parameter
<freq>
Video bandwidth (VBW)
Range/Resolution 1 Hz to 10 MHz (1 to 3 sequences), 5 kHz, OFF
Suffix code None. Value is returned in Hz units.
Example of Use
To query the VBW.
SWE:EGAT:VIEW:BAND:VID?
> 3000
[:SENSe]:SWEep:EGATe:VIEW:DETector[:FUNCtion]
NORMal|POSitive|SAMPle|NEGative|RMS
Gate View Detection Mode
Function
This command selects the waveform pattern detection mode for Gate
View.
This parameter and the one for setting the waveform pattern detection
mode for time axis measurement reference the same value.
Command
[:SENSe]:SWEep:EGATe:VIEW:DETector[:FUNCtion] <det>
Parameter
<det>
Detection mode
NORMal
Simultaneous detection of positive and negative
peaks (Default)
POSitive
Positive peak detection
NEGative
Negative peak detection
SAMPle
Sample detection
RMS
RMS detection
Details
This command is not available when Gate View is set to Off.
Example of Use
To set the detection mode to positive peak detection.
SWE:EGAT:VIEW:DET POS