Chapter
4
Native Device Message Details
4-235
4
Native Device Message Details
Details
The setting range of this function is limited depending on the settings of
other parameters. Refer to
MS2690A/MS2691A/MS2692A Signal
Analyzer Operation Manual(Spectrum Analyzer Function Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual (Spectrum
Analyzer Function Operation)
for details.
Example of Use
To set the video bandwidth of the offset automatically.
SEM:OFFS:LIST:BAND:VID:AUTO
ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON
SEM:OFFS:LIST:BAND:VID:AUTO?
> 1,1,1,1,1,1,1,1,1,1,1,1
SEM:OFFS:LIST:BAND:VID:MODE/SEM:OFFS:LIST:BAND:VID:MODE?
Spectrum Emission Mask Offset Video Bandwidth Mode
Function
This command sets the processing method for the video bandwidth(VBW)
of the offset for Spectrum Emission Mask measurement.
Command
SEM:OFFS:LIST:BAND:VID:MODE
method_1[,method_2[,method_3[,method_4[,method_5[,method
_6[,method_7[,method_8[,method_9[,method_10[,method_11[,
method_12]]]]]]]]]]]
Query
SEM:OFFS:LIST:BAND:VID:MODE?
Response
method_1,method_2,method_3,method_4,method_5,method_6,me
thod_7,method_8,method_9,method_10,method_11,method_12
Parameter
method_n
Processing method for VBW of Offset-n
VID
Normal VBW
POW
Power VBW
Example of Use
To set the method of processing the video bandwidth of the offset to
Normal VBW.
SEM:OFFS:LIST:BAND:VID:MODE
VID,VID,VID,VID,VID,VID,VID,VID,VID,VID,VID,VID
SEM:OFFS:LIST:BAND:VID:MODE?
> VID,VID,VID,VID,VID,VID,VID,VID,VID,VID,VID,VID