Chapter 2 SCPI Device Message Details
2-38
[:SENSe]:FREQuency:OFFSet <freq>
Frequency Offset
Function
This command sets the frequency offset.
Command
[:SENSe]:FREQuency:OFFSet <freq>
Parameter
<freq>
Offset
Range 100 to 100 GHz
Resolution 1 Hz
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
Default value 0 Hz
Example of Use
To set the frequency offset to 10 MHz.
FREQ:OFFS 10MHZ
[:SENSe]:FREQuency:OFFSet?
Frequency Offset Query
Function
This command queries the frequency offset.
Query
[:SENSe]:FREQuency:OFFSet?
Response
<freq>
Parameter
<freq>
Offset
Range 100 to 100 GHz
Resolution 1 Hz
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
Example of Use
To query the frequency offset.
FREQ:OFFS?
> 10000000