Chapter
Native Device Message Details
4-218
SEM:BAND:VID/SEM:BAND:VID?
Spectrum Emission Mask Video Bandwidth
Function
This command sets the video bandwidth (VBW) in measuring the
reference power for Spectrum Emission Mask measurement.
Command
SEM:BAND:VID bandwidth
SEM:BAND:VID OFF
Query
SEM:BAND:VID?
Response
bandwidth
Returns a value of Hz units, without a suffix code.
When set to Off, returns OFF.
Parameter
bandwidth
Video bandwidth (VBW)
Range/Resolution 1 Hz to 10 MHz (1-3 sequence), 5 kHz
Suffix code HZ,KHZ,KZ,MHZ,MZ,GHZ,GZ
Hz is used when omitted.
OFF
Sets VBW to Off.
Details
This command is fixed to Auto and cannot be set if the detection mode of
the measurement target is set to the following.
RMS
The setting range of this function is limited depending on the settings of
other parameters. Refer to
MS2690A/MS2691A/MS2692A Signal
Analyzer Operation Manual (Spectrum Analyzer Function Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual (Spectrum
Analyzer Function Operation)
for details.
Example of Use
To set the video bandwidth of the reference power to 3 kHz.
SEM:BAND:VID 3KHZ
SEM:BAND:VID?
> 3000