Chapter 2 SCPI Device Message Details
2-656
[:SENSe]:SEMask:OFFSet[1]:LIST:RLEVel:AUTO?
Spectrum Emission Mask Offset Reference Level Auto/Manual Query
Function
This command queries the automatic reference level setting function
On/Off state for the offset during Spectrum Emission Mask
measurement.
Query
[:SENSe]:SEMask:OFFSet[1]:LIST:RLEVel:AUTO?
Response
<switch_n>
Parameter
<switch_n>
Automatic reference level setting function
On/Off for Offset-n
1
Automatic reference level setting function is
enabled.
0
Automatic reference level setting function is
disabled.
Details
The same value as the reference level when measuring the reference
power is used in the automatic setting.
Example of Use
To query the automatic reference level setting function On/Off state for
the offset.
SEM:OFFS:LIST:RLEV:AUTO?
> 1,1,0,0,1,1,1,1,0,0,1,1