SCPI Device Message Details
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?
Frequency Step Size Query
Function
This command queries the step size of the center, start and stop
frequency.
Query
[:SENSe]:FREQuency:CENTer:STEP?
Response
<freq>
Parameter
<freq>
Step size
Range
[MS269xA]
1 Hz to 6.0 GHz (MS2690A)
1 Hz to 13.5 GHz (MS2691A)
1 Hz to 26.5 GHz (MS2692A)
[MS2830A]
1 Hz to 3.6 GHz (Option 040)
1 Hz to 6.0 GHz (Option 041)
1 Hz to 13.5 GHz (Option 043)
1 Hz to 26.5 GHz (Option 044)
1 Hz to 43 GHz (Option 045)
[MS2840A]
1 Hz to 3.6 GHz (Option 040)
1 Hz to 6.0 GHz (Option 041)
1 Hz to 26.5 GHz (Option 044)
1 Hz to 44.5 GHz (Option 046)
Resolution 1 Hz
Suffix code None. Value is returned in Hz units.
Example of Use
To query the step size.
FREQ:CENT:STEP?
> 100000