2
SCPI Device Message Details
Parameter
<switch>
Enables/disables coupling
1
The RBW, VBW, detection mode, and trace point
are coupled (shared) between the time domain
and the frequency domain.
0
The RBW, VBW, detection mode, and trace point
for the time domain are separate from those for
the frequency domain.
Example of Use
To query whether coupling is enabled.
FREQ:DOM:COUP?
> 0
[:SENSe]:FREQuency:SYNThesis[:STATe] BPHase|NORMal|FAST
Switching Speed
Function
This command sets the switching speed of frequency.
Command
[:SENSe]:FREQuency:SYNThesis[:STATe] <mode>
Parameter
<mode>
Frequency switching speed
BPHase
The operation is done so as to improve the phase
noise characteristic rather than the frequency
switching speed.
NORMal
Same as the setting of BPHase.
FAST
The operation is done so as to increase the
frequency switching speed.
Details
This command is available for MS2830A and MS2840A.
Note that because the FAST setting gives priority to the switching speed
of the local frequency, the phase noise characteristic worsens.
Example of Use
To set the frequency switching mode to the speed priority mode.
FREQ:SYNT FAST