Chapter
Native Device Message Details
4-57
4
Native Device Message Details
CALC:MARK:Y:DELT?
Marker Relative Level Query
Function
This command queries the marker level in relative value.
Query
CALC:MARK:Y:DELT? marker
Response
level
Parameter
marker
Marker Number
1
Specifies marker 1
2
Specifies marker 2
3
Specifies marker 3
4
Specifies marker 4
5
Specifies marker 5
6
Specifies marker 6
7
Specifies marker 7
8
Specifies marker 8
9
Specifies marker 9
10
Specifies marker 10
level
Relative level of zone marker
(When Scale Mode is Log)
No suffix code, in dB units, resolution: 0.001 dB
999.0 is returned at no measurement or error
(When Scale Mode is Lin)
No suffix code, no units, range 0.0000 to 10000, 0.0001 resolution
999.0 is returned at no measurement or error
Details
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Not available during Spectrum Emission Mask measurement.
Not available when Marker Mode is set other than Delta.
Example of Use
To query the relative value result of marker 1.
CALC:MARK:Y:DELT? 1
>10.00