Chapter 2 SCPI Device Message Details
2-240
:MMEMory:DELete:WAVEform:ALL [<device>]
Delete All Waveform Data Files
Function
This command deletes all Waveform data folders and all files in those
folders.
Command
:MMEMory:DELete:WAVEform:ALL [<device>]
Parameter
<device>
Drive name
A,B,D,E,F,...
Defaults to D when omitted.
Details
From the following folder in the specified drive, all data folders and all
files in those folders are deleted:
\Anritsu Corporation\Signal Analyzer\User Data\Waveform
Example of Use
To delete the Waveform data folders and all files in those folders.
MMEM:DEL:WAVE:ALL
:CALCulate:NTData[:STATe] OFF|ON|0|1
Normalize
Function
This command enables/disables the Normalize function.
Command
:CALCulate:NTData[:STATe] OFF|ON|0|1
Parameter
<switch>
Normalize function On/Off
ON|1
Enables the Normalize function
OFF|0
Disables the Normalize function (Default)
Details
This command is available only for MS2830A.
This cannot be executed under the following conditions:
When Internal Signal Generator Control On/Off is Off.
When the Math function is other than Off.
When Store Ref. is not executed.
Example of Use
To enable the Normalize function.
CALC:NTD ON