Chapter 2 SCPI Device Message Details
2-378
[:SENSe]:ACPower:OFFSet[1]:LIST[:FREQuency]?
Adjacent Channel Power Offset Frequency Query
Function
This command queries the offset frequency of Offset Channel for
Adjacent Channel Power measurement.
Query
[:SENSe]:ACPower:OFFSet[1]:LIST[:FREQuency]?
Response
<freq_1>,<freq_2>,<freq_3>
Parameter
<freq_n>
Offset frequency of the offset channel “n”
Range –1000000000 to 1000000000 Hz
Resolution 1 Hz
Suffix code None. Value is returned in Hz units.
Example of Use
To query the offset frequency of the offset channel.
ACP:OFFS:LIST?
> 30000,50000,50000
Related command
This command has the same function as the following command.
:CALCulate:ACPower:OFFSet[1]:LIST[:FREQuency]?
:CALCulate:ACPower:OFFSet[1]:LIST[:FREQuency]?
Adjacent Channel Power Offset Frequency Query
Function
This command queries the offset frequency of the offset channel for
Adjacent Channel Power measurement.
Refer to
[:SENSe]:ACPower:OFFSet[1]:LIST[:FREQuency]?.
Related command
This command has the same function as the following command.
[:SENSe]:ACPower:OFFSet[1]:LIST[:FREQuency]?