2.9 ACP
2-357
2
SCPI Device Message Details
[:SENSe]:ACPower:CARRier[1]:RCARrier:METHod
STOTal|CTOTal|BSIDes|CSELect
Adjacent Channel Power Reference
Function
This command sets the reference of the relative level display for ACP
measurement.
Command
[:SENSe]:ACPower:CARRier[1]:RCARrier:METHod <method>
Parameter
<method>
Reference method for the relative level display
for ACP measurement.
STOTal
Sets the integral power on the entire screen as a
reference (Span Total method).
CTOTal
Sets the total value of all carrier power as a
reference (Carrier Total method)(Default).
BSIDes
Carrier power of the largest carrier number is
used as a reference for the upper offset, while
the carrier power of the smallest carrier number
is used as reference (Both Sides of Carriers
method).
CSELect
Uses the carrier selected in Carrier Select as a
reference.
Example of Use
To set the ACP measurement method to Carrier Total method.
ACP:CARR:RCAR:METH CTOT
Related command
This command has the same function as the following command.
:CALCulate:ACPower:CARRier[1]:RCARrier:METHod