Chapter 2 SCPI Device Message Details
2-160
:CALCulate:BANDwidth|:BWIDth:NDB:STATe OFF|ON|0|1
N dB Bandwidth
Function
This command enables/disables the N dB Bandwidth measurement
function.
Command
:CALCulate:BANDwidth|:BWIDth:NDB:STATe <switch>
Parameter
<switch>
N dB Bandwidth measurement function On/Off
ON|1
Enables N dB Bandwidth measurement function
OFF|0
Disables N dB Bandwidth measurement
function (Default)
Details
This command is available only for MS2830A.
This cannot be executed under the following conditions:
• In Time Domain
• When Scale Mode is Linear
• When Gate View is On
• When the Measure function is On
Example of Use
To enable the N dB Bandwidth measurement function.
CALC:BAND:NDB:STAT ON
:CALCulate:BANDwidth|:BWIDth:NDB:STATe?
N dB Bandwidth Query
Function
This command queries the On/Off state of the N dB Bandwidth
measurement function.
Query
:CALCulate:BANDwidth|:BWIDth:NDB:STATe?
Response
<switch>
Parameter
<switch>
N dB Bandwidth measurement function On/Off
1
N dB Bandwidth measurement function is enabled.
0
N dB Bandwidth measurement function is disabled.
Details
This command is available only for MS2830A.
Example of Use
To query the On/Off state of the N dB Bandwidth measurement function.
CALC:BAND:NDB:STAT?
> 1