2.14 Spurious Emission Measurement Function
2-743
2
SCPI Device Message Details
:DISPlay:SPURious:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:RLEVel?
Spurious Emission Reference Level Query
Function
This command queries the reference level of each segment.
Query
:DISPlay:SPURious:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:RLEV
el?
Response
<real_n>
Parameter
<real_n>
Reference level of Segment n
Range Value equivalent to120 to +50 dBm
Resolution 0.01 dB
Suffix Code None. Value is returned in dBm units.
Example of Use
To query the reference level of each segment.
DISP:SPUR:VIEW:WIND:TRAC:Y:RLEV?
>
0.00,0.00,0.00,3.00,0.00,0.00,0.00,8.00,0.00,0.00,0.00,0