Chapter
Native Device Message Details
4-262
SIGID/SIGID?
Signal Identifier
Function
This command enables/disables signal identifier operation, which
identifies measured signal and image signal when external mixer is used.
Command
SIGID switch
Query
SIGID?
Response
switch
Parameter
switch
Signal ID
1
Selects signal ID.
0
Does not select signal ID.
Details
This function is available when Option 044/045 is installed for MS2830A.
This function is available when Option 044/046 is installed for MS2840A.
Signal ID cannot be set to On when the PS function is On.
Signal ID cannot be set to On when the Measure function is On
Signal ID cannot be set to On when the Math function is other than Off.
Example of Use
To sweep while identifying measured signal and image signal.
SIGID 1
SIGID?
> 1