Native Device Message Details
4-100
FREQ:DOM:COUP/FREQ:DOM:COUP?
Couple Time/Frequency Domain
Function
Sets whether to couple the time domain parameters and frequency
domain parameters.
Command
FREQ:DOM:COUP switch_com
Query
FREQ:DOM:COUP?
Response
switch_res
Parameter
switch_com
Enables/disables coupling
ON|1
Couples (shares) the RBW, VBW, detection mode,
and trace point between the time domain and
the frequency domain.
OFF|0
Separates the RBW, VBW, detection mode, and
trace point for the time domain from those for
the frequency domain.
switch_res
Coupling enabled/disabled state
1
The RBW, VBW, detection mode, and trace point
are coupled (shared) between the time domain
and the frequency domain.
0
The RBW, VBW, detection mode, and trace point
for the time domain are separate from those for
the frequency domain.
Example of Use
To disable coupling.
FREQ:DOM:COUP OFF
FREQ:DOM:COUP?
> 0