2.14 Spurious Emission Measurement Function
2-775
2
SCPI Device Message Details
:CALCulate:SPURious[:RANGe][:LIST]:LIMit:ABSolute[:UPPer]:DATA:STOP:
AUTO?
Spurious Emission Limit Stop Level Auto/Manual Query
Function
This command queries Auto/Manual of the absolute level limit of the stop
frequency in each segment for the Spurious Emission measurement.
Query
:CALCulate:SPURious[:RANGe][:LIST]:LIMit:ABSolute[:UPPer
]:DATA:STOP:AUTO?
Response
<switch_n>
Parameter
<switch_n>
Stop frequency of Segment n Auto/Manual
ON|1
Stop frequency is set to Auto.
OFF|0
Stop frequency is set to Manual.
Example of Use
To query Auto/Manual of the absolute level limit of the stop frequency in
each segment.
CALC:SPUR:LIM:ABS:DATA:STOP:AUTO?
> 1,1,1,1,0,0,0,1,1,1,1,1,1,1,1,1,1,1,1,1