2
SCPI Device Message Details
[:SENSe]:BANDwidth|:BWIDth[:RESolution]:AUTO ON|OFF|1|0
Resolution Bandwidth Auto/Manual
Function
This command enables/disables the resolution bandwidth (RBW)
automatic setting function. When Couple Time/Freq. Domain is set to Off,
each parameter is specified in accordance with the already-specified
frequency axis or time axis measurement.
Command
[:SENSe]:BANDwidth|:BWIDth[:RESolution]:AUTO <switch>
Parameter
MS269xA, MS2830A, MS2840A RBW Mode Normal
<switch>
Automatic resolution bandwidth setting
function On/Off state
0|OFF
Disables the automatic resolution bandwidth
function.
1|ON
Enables the automatic resolution bandwidth
setting function (Default).
MS2830A, MS2840A RBW Mode CISPR
(Only with MS2830A-016/116, MS2840A-016/116)
<switch>
Automatic setting (for CISPR RBW)
0|OFF
Automatic setting is set to Off.
1|ON
Automatic setting is set to On (Default).
Details
MS269xA, MS2830A, MS2840A RBW Mode Normal
This command is not available during the Spectrum Emission Mask
measurement or the Spurious Emission measurement.
MS2830A, MS2840A RBW Mode CISPR
Not available when the Measure function is set to On.
Not available when the Gate View function is set to On.
Example of Use
To enable the automatic resolution bandwidth setting function.
BAND:AUTO ON