EasyManua.ls Logo

Anritsu MS2840A - :CALCulate:LLINe[1]|2|3|4|5|6:DATA <x-axis_1>,<ampl_1>,<connected_1>[,<x-axis_2>,<ampl_2>,<connected_2>],,,[,<x-axis_n>,<ampl_n>,<connected_n>]

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.6 Trace
2-195
2
SCPI Device Message Details
:CALCulate:LLINe[1]|2|3|4|5|6:DATA
<x-axis_1>,<ampl_1>,<connected_1>[,<x-axis_2>,<ampl_2>,<connected_2>]
,,,[,<x-axis_n>,<ampl_n>,<connected_n>]
Limit Edit
Function
This command sets the
Frequency
,
Amplitude
, and
Connected to Previous
Pt
parameters for Limit Points on the specified Limit Line.
Command
:CALCulate:LLINe[1]|2|3|4|5|6:DATA
<x-axis_1>,<ampl_1>,<connected_1>[,<x-axis_2>,<ampl_2>,<
connected_2>],,,[,<x-axis_n>,<ampl_n>,<connected_n>]
Parameter
<x-axis_n>
Frequency Offset
Range 0 to 325 GHz (Fixed/Absolute)
100 GHz to 100 GHz (Relative)
Default Start Frequency (Fixed/Absolute)
0 Hz (Relative)
Resolution 1 Hz
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
<
ampl
_n>
Amplitude
Range
300 to dBm (Fixed/Absolute)
300 to 300 dB (Relative)
Default Reference Level (Fixed/Absolute)
0 dB (Relative)
Resolution 0.01 dB
Suffix code None
<connected_n>
Connected to Previous point
OFF|0
Disabled (No Connected)
ON|1
Enabled (Connected) (Default)
Details
Up to 100 points can be set.
When you try to set more than 100 points, the error message
“Insufficient data” is displayed.
When a Limit points Frequency value is higher than the other Limit
points value and this relationship reverses, the order of Limit points
changes.
Example of Use
To set on Limit line1 to 1 GHz, 20 dBm, and No Connected for Point1,
and set to 2 GHz, 30 dBm, and Connected for Point2.
CALC:LLIN1:DATA 1000000000,-20,0,2000000000,-30,1