EasyManua.ls Logo

Anritsu MS2840A - :CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10:MODE?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 SCPI Device Message Details
2-104
:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10:MODE?
Marker Mode Query
Function
This command queries the marker mode.
Query
:CALCulate:MARKer[n]:MODE?
Response
<mode>
Parameter
<n>
Marker number
1
Marker 1
2
Marker 2
3
Marker 3
4
Marker 4
5
Marker 5
6
Marker 6
7
Marker 7
8
Marker 8
9
Marker 9
10
Marker 10
When omitted:
Marker 1
<mode>
Marker mode
NORM
Normal marker
DELT
Delta marker
FIX
Fixed marker
OFF
Not displaying any marker
Example of Use
To query the mode of the marker.
CALC:MARK:MODE?
> DELT
Related command
This command has the same function as the following commands.
:CALCulate:ACPower:MARKer[1]|2|3|4|5|6|7|8|9|10:MODE?
:CALCulate:CHPower:MARKer[1]|2|3|4|5|6|7|8|9|10:MODE?
:CALCulate:OBWidth:MARKer[1]|2|3|4|5|6|7|8|9|10:MODE?
:CALCulate:BPOWer|:TXPower:MARKer[1]|2|3|4|5|6|7|8|9|10:
MODE?

Table of Contents

Other manuals for Anritsu MS2840A

Related product manuals