Chapter 2 SCPI Device Message Details
2-80
:CALCulate:BANDwidth|:BWIDth[:RESolution] <freq>
Resolution Bandwidth
Function
This command sets the resolution bandwidth (RBW). When Couple
Time/Freq. Domain is set to Off, each parameter is specified in
accordance with the already-specified frequency axis or time axis
measurement.
Command
:CALCulate:BANDwidth|:BWIDth[:RESolution] <freq>
Parameter
<freq>
Resolution bandwidth (RBW)
[MS269xA]
Range/Resolution 30 Hz to 31.25 MHz
However, the settable value is 30 Hz to 3 MHz
(for 1 to 3 sequences), 50 kHz, 5 MHz, 10 MHz,
or 20 MHz, or 31.25 MHz.
[
MS2830A],
[
MS2840A]
RBW Mode Normal
Range/Resolution 1 Hz to 31.25 MHz
However, the settable value is 1 Hz to 3 MHz
(for 1 to 3 sequences), 500 Hz, 50 kHz, 2 MHz, 5
MHz, 10 MHz, 20 MHz, or 31.25 MHz.
RBW Mode CISPR (Only with MS2830A-016/116, MS2840A-016/116)
Range/Resolution The settable value is 200 Hz, 9 kHz, 120 kHz, or
1 MHz.
Default Value set in CISPR RBW Auto
[Common]
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
Default Value set in RBW Auto