2
SCPI Device Message Details
[:SENSe]:ACPower:ADVanced:FILTer[:RRC]:ALPHa <real>
Adjacent Channel Power Offset Rolloff Ratio
Function
This command sets the roll-off rate of the Adjacent Channel Power
measurement offset channel filter.
Command
[:SENSe]:ACPower:ADVanced:FILTer[:RRC]:ALPHa
<real_1>,<real_2>,<real_3>,<real_4>,<real_5>,<real_6>,<r
eal_7>,<real_8>
Parameter
<real_n>
Offset_n filter roll-off rate
Range 0.01 to 1.00
Resolution 0.01
Suffix code None
Details
This function can be set when the following trace is active.
•
Spectrum
This function is enabled when the ACP measurement offset channel filter
type (ACP Offset Filter Type) is one of the following:
•
Nyquist
•
Root Nyquist
To query the measurement results after executing this command, use the
*WAI command to control synchronization.
Example of Use
To set the roll-off rate of the ACP measurement offset channel filter to
0.22 and query the result.
ACP:ADV:FILT:ALPH 0.22,0.22,0.22,0.22,0.22,0.22,0.22,0.22
*WAI
FETC:ACP?