EasyManua.ls Logo

Anritsu MS2840A - SEM:OFFS:LIST:BAND:AUTO;SEM:OFFS:LIST:BAND:AUTO?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter
Native Device Message Details
4-229
4
Native Device Message Details
Example of Use
To set the resolution bandwidth of the offset.
SEM:OFFS:LIST:BAND
30KHZ,30KHZ,30KHZ,1MHZ,1MHZ,1MHZ,30KHZ,30KHZ,30KHZ,1MHZ,
1MHZ,1MHZ
SEM:OFFS:LIST:BAND?
>
30000,30000,30000,1000000,1000000,1000000,30000,30000,30
000,1000000,1000000,1000000
SEM:OFFS:LIST:BAND:AUTO/SEM:OFFS:LIST:BAND:AUTO?
Spectrum Emission Mask Offset Resolution Bandwidth Auto/Manual
Function
This command sets the resolution bandwidth (RBW) of the offset for
Spectrum Emission Mask measurement automatically.
Command
SEM:OFFS:LIST:BAND:AUTO
switch_1_com[,switch_2_com[,switch_3_com[,switch_4_com[,
switch_5_com[,switch_6_com[,switch_7_com[,switch_8_com[,
switch_9_com[,switch_10_com[,switch_11_com[,switch_12_co
m]]]]]]]]]]]
Query
SEM:OFFS:LIST:BAND:AUTO?
Response
switch_1_res,switch_2_res,switch_3_res,switch_4_res,swit
ch_5_res,switch_6_res,switch_7_res,switch_8_res,switch_9
_res,switch_10_res,switch_11_res,switch_12_res
Parameter
switch_n_com
Automatic setting function of RBW in measuring
Offset-n
ON|1
Sets the automatic setting function to On.
OFF|0
Sets the automatic setting function to Off.
switch_n_res
Automatic setting function of RBW in measuring
Offset-n
1
Automatic setting function is On.
0
Automatic setting function is Off.