EasyManua.ls Logo

Anritsu MS2840A - [:SENSe]:SWEep[:TYPE][:AUTO]:RULes:FFT:WIDTh <freq>

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.7 Sweep/Trigger/Gate
2-263
2
SCPI Device Message Details
[:SENSe]:SWEep[:TYPE][:AUTO]:RULes:FFT:WIDTh <freq>
Sweep Type Select Rules FFT Width
Function
This command sets the FFT width when FFT Priority is selected, for the
sweep/FFT Switching rules during measurement.
Command
[:SENSe]:SWEep[:TYPE][:AUTO]:RULes:FFT:WIDTh <freq>
Parameter
<freq>
FFT width
Range 40 kHz or 2 MHz
Resolution 1 Hz
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
Default 40 kHz
Details
This command is not available for MS269x Series.
When a value other than 40 kHz or 2 MHz is set, the value is rounded off
to the smaller value and set.
This command is available only when FFT Priority is selected for Auto
Sweep Type Rules.
The specified width is used as the FFT width on a priority basis, but it is
automatically changed by the SPAN/RBW/VBW measurement
conditions.
Example of Use
To set FFT width to 40 kHz.
SWE:RUL:FFT:WIDT 40KHZ

Table of Contents

Other manuals for Anritsu MS2840A

Related product manuals