Chapter
Native Device Message Details
4-255
4
Native Device Message Details
SEM:SWE:POIN/SEM:SWE:POIN?
Spectrum Emission Mask Trace Point
Function
This command selects the frequency display point in measuring the
reference power for Spectrum Emission Mask measurement.
Command
SEM:SWE:POIN integer
Query
SEM:SWE:POIN?
Response
integer
Parameter
integer
Number of frequency point
Range 11 to 30001
Resolution Any of the following values can be selected:
11, 21, 41, 51, 101, 201, 251, 401, 501, 1001,
2001, 5001, 10001, 30001 (MS269xA Only)
Suffix code None
Example of Use
To set the frequency display point number in measuring the reference
power to 2001.
SEM:SWE:POIN 2001
SEM:SWE:POIN?
> 2001