SCPI Device Message Details
S is used when omitted.
Details
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Not available during Spectrum Emission Mask measurement.
When Internal Signal Generator Control On/Off is On, the center
frequency range of zone marker depends on the frequency range of Signal
Generator.
Example of Use
To move the center frequency of the zone marker to 100 MHz.
CALC:MARK:X 100MHZ
Related command
This command has the same function as the following commands.
:CALCulate:ACPower:MARKer[1]|2|3|4|5|6|7|8|9|10:X
:CALCulate:CHPower:MARKer[1]|2|3|4|5|6|7|8|9|10:X
:CALCulate:OBWidth:MARKer[1]|2|3|4|5|6|7|8|9|10:X
:CALCulate:BPOWer|:TXPower:MARKer[1]|2|3|4|5|6|7|8|9|10:X
:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10:X?
Zone Marker Frequency (Time) Query
Function
This command queries the center of the zone marker.
Query
:CALCulate:MARKer[n]:X?
Response
<freq>
<time>
Parameter
<n>
Marker number
1
Marker 1
2
Marker 2
3
Marker 3
4
Marker 4
5
Marker 5
6
Marker 6
7
Marker 7
8
Marker 8
9
Marker 9
10
Marker 10