Native Device Message Details
4-326
SWE:EGAT:VIEW:BAND:VID/SWE:EGAT:VIEW:BAND:VID?
Gate View Video Bandwidth
Function
This command sets the video bandwidth (VBW) for Gate View.
This parameter and the one for setting the VBW for time axis
measurement reference the same value.
Command
SWE:EGAT:VIEW:BAND:VID freq
Query
SWE:EGAT:VIEW:BAND:VID?
Response
freq
Parameter
freq
Video bandwidth (VBW)
Range/Resolution 1 Hz to 10 MHz (1-3 sequence), 5 kHz, OFF
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
Details
The automatic video bandwidth setting function is disabled when the
video bandwidth (VBW) is changed.
This command is not available when Gate View is set to Off.
This command is not available when Gate View Detection is set to RMS.
Example of Use
To set the VBW to 3 kHz.
SWE:EGAT:VIEW:BAND:VID 3KHZ
SWE:EGAT:VIEW:BAND:VID?
> 3000