2.4 Marker
2-145
2
SCPI Device Message Details
:CALCulate:MARKer:REFerence:X:POSition?
Reference Marker Position Query
Function
This command queries the position of the marker specified by Relative To
of the active marker by the display points from the left edge of the
screen.
Query
:CALCulate:MARKer:REFerence:X:POSition?
Response
<integer>
Parameter
<integer>
Marker position set in Relative To (Number of
display points from the left edge of the screen)
Range 0 to 10, 20, 40, 50, 100, 200, 250, 400, 500, 1000,
2000, 5000, 10000, 30000 (MS269xA Only)
(The upper limit varies depending on the
number of the trace display points.)
Resolution 1
Details
999.0 is returned when Marker Mode is set to other than Delta.
This command queries the result in absolute values even if the marker
mode specified by Relative To is set to Delta.
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Example of Use
To query the position of the marker point set in Relative To.
CALC:MARK:REF:X:POS?
> 123