Chapter
4
Native Device Message Details
4-163
4
Native Device Message Details
MKLTYPE/MKLTYPE?
Marker Result
Function
This command sets the type of the marker value.
Command
MKLTYPE type_com
Query
MKLTYPE?
Response
type_res
Parameter
type_com
Type of Marker value
INT
Integral power in the zone
TOTAL
Same as above
DENS
Power density in the zone
AVG
Same as above
PEAK
Peak level in the zone
type_res
Type of Marker value
TOTAL
Integral power in the zone
AVG
Power density in the zone
PEAK
Peak level in the zone
Details
When Scale Mode is Lin, Integration and Density cannot be set.
Integration cannot be set in the time domain mode.
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Not available during Spectrum Emission Mask measurement.
When N dB Bandwidth Measurement is On, the Type of Marker display
value is set to the peak level in the zone.
Example of Use
To set the marker value to the power density within the zone band.
MKLTYPE AVG
MKLTYPE?
> AVG