2.7 Sweep/Trigger/Gate
2-257
2
SCPI Device Message Details
[:SENSe]:SWEep:TIME:AUTO?
Sweep Time Auto/Manual Query
Function
This command queries the On/Off state of the automatic sweep time
setting function. Each parameter is specified in accordance with the
already-specified frequency axis or time axis measurement.
Query
[:SENSe]:SWEep:TIME:AUTO?
Response
<switch>
Parameter
<switch>
Automatic sweep time setting function On/Off
0
Automatic sweep time setting function is
disabled.
1
Automatic sweep time setting function is
enabled (Default).
Example of Use
To query the On/Off state of the automatic sweep time setting function.
SWE:TIME:AUTO?
> 1
Related command
This command has the same function as the following commands.
[:SENSe]:ACPower:SWEep:TIME:AUTO?
[:SENSe]:CHPower:SWEep:TIME:AUTO?
[:SENSe]:OBWidth:SWEep:TIME:AUTO?
[:SENSe]:SWEep:TIME:AUTO:MODE NORMal|FAST
Auto Sweep Time Mode
Function
This command sets the sweep mode (Fast or Normal) when the
automatic sweep time setting function is enabled.
Command
[:SENSe]:SWEep:TIME:AUTO:MODE <mode>