Chapter 2 SCPI Device Message Details
2-662
:CALCulate:SEMask:OFFSet[1]:LIST:STARt:ABSolute[1]|2?
Spectrum Emission Mask Offset Start Frequency Absolute Limit Level Query
Function
This command queries the absolute level upper limit of the offset start
frequency for Spectrum Emission Mask measurement.
Query
:CALCulate:SEMask:OFFSet[1]:LIST:STARt:ABSolute[1]|2?
Response
<ampl_n>
Parameter
<ampl_n>
Absolute level upper limit of Offset-n start
frequency
Range 200 to +50 dBm
Resolution 0.01 dB
Suffix code None. Value is returned in dBm units.
Details
ABSolute1 sets the absolute level upper limit 1, and ABSolute2 sets the
absolute level upper limit 2.
Example of Use
To query the absolute level upper limit 2 of the start frequency of the
offset.
CALC:SEM:OFFS:LIST:STAR:ABS2?
>
0.00,-2.00,-2.00,0.00,0.00,-10.00,0.00,-2.00,-2.00,0.00,
0.00,-10.00
Related command
This command has the same function as the following command.
[:SENSe]:SEMask:OFFSet[1]:LIST:STARt:ABSolute[1]|2?