Chapter 2 SCPI Device Message Detail
2-858
:MEASure:BATCh:ACP[n]? <filename>[,<device>]
Adjacent Channel Power Batch Measure
Function
This command executes ACP measurement and outputs the result after
the parameters described in the specified parameter list file have been
set.
Query
:MEASure:BATCh:ACP[n]? <filename>[,<device>]
Response
The same value as for
:MEASure:ACP[n]?
is returned.
Refer to
:MEASure:ACP[n]?
Parameter
<filename>
Parameter list file
Specify with any character string enclosed by
double quotes (" ") or single quotes (' ').
<device>
Drive name
A,B,D,E,F,...
D drive is used when omitted.
Details
This command executes ACP measurement and outputs the result after
the parameters described in the specified parameter list have been set.
The return values of this function vary depending on the result mode.
(cf.
:SYSTem:RESut:MODE
)
Place the parameter list files in the following folder on the specified
drive.
<device>:\Anritsu Corporation\Signal Analyzer\User Data\Batch
If the parameter list file has been changed, the changes must be applied
with the
:MMEMory:RELoad:BATCh
command.
(cf.
:MMEMory:RELoad:BATCh
)
Example of Use
To acquire the measurement result of ACP measurement using the
MyParam.xls parameter list file.
MEAS:BATC:ACP? "MyParam"
>
0.0,-72.130,0.0,-72.130,-1.270,-73.400,-0.570,-72.700,-0
.780,-72.910,-1.030,-73.160,-999.0,-999.0,-999.0,-999.0