EasyManua.ls Logo

Anritsu MS2840A - INIT:SPUR:PAUS:CONT

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter
Native Device Message Details
4-113
4
Native Device Message Details
INIT:SPUR:PAUS:CONT
Spurious Emission Continue
Function
Pause is cancelled and measurement will continue when this command is
sent while it is in remote control state and paused.
Command
INIT:SPUR:PAUS:CONT
Details
It pauses before the pertinent segment is swept when Pause before
Sweep is set to On. This command is used to cancel the pause and
continue the measurement.
This command can only be used while it is in remote control state and
paused.
Pause is cancelled and measurement stops when this command is sent
while it is in local status. Press the F1 [Continue] key to continue the
measurement while it is in local status.
To query the measurement results after executing this command,
perform the synchronization control using the “*WAI” command.
Example of Use
Measurement method when pausing twice
INIT:SPUR
Measure until paused
*WAI
Wait until paused
INIT:SPUR:PAUS:CONT
Cancel the pause, and continue the
measurement
*WAI
Wait until paused
INIT:SPUR:PAUS:CONT
Cancel the pause, and continue the
measurement
*WAI
Wait until the measurement is
completed
FETC:SPUR? 1
Read the measurement results
> 1,1,1,9282.00,-84.38,71.38,...