EasyManua.ls Logo

Anritsu MS2840A - [:SENSe]:ACPower:OFFSet[1]:LIST[:FREQuency] <freq>,<freq>,<freq>; :CALCulate:ACPower:OFFSet[1]:LIST[:FREQuency] <freq>,<freq>,<freq>

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.9 ACP
2-377
2
SCPI Device Message Details
[:SENSe]:ACPower:OFFSet[1]:LIST[:FREQuency] <freq>,<freq>,<freq>
Adjacent Channel Power Offset Frequency
Function
This command sets the offset frequency of Offset Channel for Adjacent
Channel Power measurement.
Command
[:SENSe]:ACPower:OFFSet[1]:LIST[:FREQuency]
<freq_1>,<freq_2>,<freq_3>
Parameter
<freq_n>
Offset frequency of offset channel “n”.
Range 1000000000 to 1000000000 Hz
Resolution 1 Hz
Suffix code HZ,KHZ,KZ,MHZ,MZ,GHZ,GZ
Hz is used when omitted.
Default
freq-1
5 MHz
freq-2
10 MHz
freq-3
15 MHz
Example of Use
To set the offset frequency of the offset channel.
ACP:OFFS:LIST 30KHZ,50KHZ,50KHZ
Related command
This command has the same function as the following command.
:CALCulate:ACPower:OFFSet[1]:LIST[:FREQuency]
:CALCulate:ACPower:OFFSet[1]:LIST[:FREQuency] <freq>,<freq>,<freq>
Adjacent Channel Power Offset Frequency
Function
This command sets the offset frequency of Offset Channel for Adjacent
Channel Power measurement.
Refer to
[:SENSe]:ACPower:OFFSet[1]:LIST[:FREQuency].
Related command
This command has the same function as the following command.
[:SENSe]:ACPower:OFFSet[1]:LIST[:FREQuency]

Table of Contents