2.7 Sweep/Trigger/Gate
2-277
2
SCPI Device Message Details
:TRIGger[:SEQuence]:WIF|:RFBurst:LEVel:ABSolute <level>
Wide IF Trigger Level
Function
This command sets the threshold of the level where trigger sweep starts
in Wide IF Video trigger.
Command
:TRIGger[:SEQuence]:WIF|:RFBurst:LEVel:ABSolute <level>
Parameter
<level>
Threshold value of level where trigger sweep
starts
Range –60 to 50 dBm
Resolution 1 dB
Initial
Default –20 dBm
Example of Use
To set the threshold of the level where trigger sweep starts to 10 dBm.
TRIG:WIF:LEV:ABS 10
:TRIGger[:SEQuence]:WIF|:RFBurst:LEVel:ABSolute?
Wide IF Trigger Level Query
Function
This command queries the threshold of the level where trigger sweep
starts in Wide IF Video trigger.
Query
:TRIGger[:SEQuence]:WIF|:RFBurst:LEVel:ABSolute?
Response
<level>
Parameter
<level>
Threshold of the level where trigger sweep
Range –60 to 50 dBm
Resolution 1 dB
Suffix code None. Value is returned in dBm units.
Example of Use
To query the threshold of the level where trigger sweep starts.
TRIG:WIF:LEV:ABS?
> 10