Chapter
4
Native Device Message Details
4-81
4
Native Device Message Details
DISP:SPUR:VIEW:WIND:TRAC:Y:RLEV/DISP:SPUR:VIEW:WIND:TRAC:Y:R
LEV?
Spurious Emission Reference Level
Function
This command sets the reference level of each segment.
Command
DISP:SPUR:VIEW:WIND:TRAC:Y:RLEV
real_1,real_2,real_3,real_4,real_5,real_6,real_7,real_8,
real_9,real_10,real_11,real_12,real_13,real_14,real_15,r
eal_16,real_17,real_18,real_19,real_20
Query
DISP:SPUR:VIEW:WIND:TRAC:Y:RLEV?
Response
real_1,real_2,real_3,real_4,real_5,real_6,real_7,real_8,
real_9,real_10,real_11,real_12,real_13,real_14,real_15,r
eal_16,real_17,real_18,real_19,real_20
Value is returned in dBm units, without suffix code.
Parameter
real_n
Reference level of Segment n
Range Value equivalent to –120 to +50 dBm
Resolution 0.01 dB
Suffix code DBM,DM
dBm is used when omitted.
Default 0 dBm
Example of Use
To set the reference level of each segment.
DISP:SPUR:VIEW:WIND:TRAC:Y:RLEV
0DBM,0,0,3,0,0,0,8,0,0,0,0,0,0,0,0,0,0,0,0
DISP:SPUR:VIEW:WIND:TRAC:Y:RLEV?
>
0.00,0.00,0.00,3.00,0.00,00.0,0.00,8.00,0.00,0.00,0.00,0
.00,0.00,0.00,0.00,0.00,0.00,0.00,0.00,0.00