Chapter
4
Native Device Message Details
4-86
Range 0 to 255
Example of Use
To enable the measurement end (End of Measure) status.
ESE2 32
ESE2?
> 1
ESR2?
END Event Status Register Query
Function
This command queries the END event status register value. Clears the
END event status register after readout.
Query
ESR2?
Response
n
Parameter
n
END event status register
Value = bit0 + bit1 + bit2 + bit3 + bit4 + bit5 + bit6 +
bit7
bit0: 2
0
= 1 End of sweep
bit1: 2
1
= 2 (Not used)
bit2: 2
2
= 4 (Not used)
bit3: 2
3
= 8 (Not used)
bit4: 2
4
= 16 End of Average
bit5: 2
5
= 32 End of Measure
bit6: 2
6
= 64 End of Max/Min Hold
bit7: 2
7
= 128 (Not used)
Range 0 to 255
Example of Use
To query the END event status register value. (End of sweep)
ESR2?
> 1