2.14 Spurious Emission Measurement Function
2-737
2
SCPI Device Message Details
[:SENSe]:SPURious[:RANGe][:LIST]:FREQuency:STARt?
Spurious Emission Start Frequency Query
Function
This command queries the start frequency of each segment.
Query
[:SENSe]:SPURious[:RANGe][:LIST]:FREQuency:STARt?
Response
<freq_n>
Parameter
<freq_n>
Start frequency of segment n
[MS269xA]
–
100
MHz to 6.0499997 GHz (MS2690A)
–
100
MHz to 13.5999997 GHz (MS2691A)
–
100
MHz to 26.5999997 GHz (MS2692A)
[MS2830A]
–
100
MHz to 3.6999997 GHz (Option 040)
–
100
MHz to 6.0999997 GHz (Option 041)
–
100
MHz to 13.5999997 GHz (Option 043)
–
100 MHz to 26.5999997 GHz (Option 044)
–
100 MHz to 43.0999997 GHz (Option 045)
[MS2840A]
–
100
MHz to 3.6999997 GHz (Option 040)
–
100
MHz to 6.0999997 GHz (Option 041)
–
100 MHz to 26.9999997 GHz (Option 044)
–
100 MHz to 44.9999997 GHz (Option 046)
Resolution 1 Hz
Suffix code None. Value is returned in Hz units.
Example of Use
To query the start frequency of each segment.
SPUR:FREQ:STAR?
>
9000,100000,40000000,11000000000,1000000000,1000000000,1
000000000,1000000000,1000000000,1000000000,1000000000,10
00000000,1000000000,1000000000,1000000000,1000000000,100
0000000,1000000000,1000000000,1000000000