EasyManua.ls Logo

Anritsu MS2840A - :CALCulate:SEMask:OFFSet[1]:LIST:STARt:RCARrier <ampl_1>[,<ampl_2>[,<ampl_3>[,<ampl_4>[,<ampl_5>[,<ampl_6>[,<ampl_7>[,<ampl_8>[,<ampl_9>[,<ampl_10>[,<ampl_11>[,<ampl_12>]]]]]]]]]]]

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.13 Spectrum Emission Mask
2-669
2
SCPI Device Message Details
:CALCulate:SEMask:OFFSet[1]:LIST:STARt:RCARrier
<ampl_1>[,<ampl_2>[,<ampl_3>[,<ampl_4>[,<ampl_5>[,<ampl_6>[,<ampl_7
>[,<ampl_8>[,<ampl_9>[,<ampl_10>[,<ampl_11>[,<ampl_12>]]]]]]]]]]]
Spectrum Emission Mask Offset Start Frequency Limit Level
Function
This command sets the relative level upper limit of the offset start
frequency for Spectrum Emission Mask measurement.
Command
:CALCulate:SEMask:OFFSet[1]:LIST:STARt:RCARrier <ampl_n>
Parameter
<ampl_n>
Relative level upper limit of Offset-n start
frequency
Range 200 to +50 dB
Resolution 0.01 dB
Suffix code DB
dB is set when omitted.
Example of Use
To set the relative level upper limit of the start frequency of the offset.
CALC:SEM:OFFS:LIST:STAR:RCAR 0,0,0,0,0,0,0,0,0,0,0,0
Related command
This command has the same function as the following command.
[:SENSe]:SEMask:OFFSet[1]:LIST:STARt:RCARrier

Table of Contents

Other manuals for Anritsu MS2840A

Related product manuals